• Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
  • Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
  • Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
  • Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
  • Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
  • Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter

Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter

Type: Multi-Parameter Water Quality Meter
Display: Digital
Application: Chemicals
Portable: Portable
Off Mode: Automatic
pH Range: -2.000~20.000pH
Customization:
Manufacturer/Factory & Trading Company

360° Virtual Tour

Diamond Member Since 2011

Suppliers with verified business licenses

Rating: 5.0/5
Shandong, China
Importers and Exporters
The supplier has import and export rights
Years of Export Experience
The export experience of the supplier is more than 10 years
Experienced Team
The supplier has 36 foreign trading staff(s) and 33 staff(s) with over 6 years of overseas trading experience
Exhibition Experience
The supplier had participated in offline trade shows, you can check the Audit Report for more information
to see all verified strength labels (34)

Basic Info.

Model NO.
PH900P
pH Accuracy
+-0.002pH
Resolution
0.001, 0.01, 0.1pH, Selectable
Calibration Points
up to 5 Points
Mv Range
-1999.9~1999.9mv
TDS Factor
0.1~1.0 (Default 0.5)
Salinity Range
0~10ppt (Max.80ppt)
Transport Package
Standard Export Package
Trademark
BIOBASE
Origin
China
Production Capacity
10000 Units Per Month

Product Description

pH Mode:
Up to 5 points push-button calibration with auto-buffer recognition.
selectable pH buffer standards, resolution and temperature units.
Automatically calculates and  displays electrode slope  that helps youdecide whether to replace pH electrode.

ORP Mode:
Single point  offset calibration allows user  adjusts displayed value toa known standard.
Absolute or Relative  mV measurements provides accurate readings.

Ion Mode:
Standard BNC connector can connects many Ion selective Electrode.
2 to 5 points push-button calibration from low to high concentrations.
selectable multiple concentration units, including ppm, mg/L, mol/L.
Direct readout ion concentration simplifies measurement process.
mV measurement allows user to check performance of ion electrode.

Conductivity/TDS/Salinity Mode:
Up to 5-point push-button calibration allows user to uses customizedcalibration solutions.
Adjustable TDS factors, cell constants, temperature coefficients andnormalization temperature.
selectable linear or pure water compensation modes.
In salinity measurement modes, selectable practical salinity (PSU) orseawater measurement standard (ppt).  

Dissolved Oxygen Mode:
1 or 2 points auto-calibration (100% in saturated air or air-saturatedwater, 0% in zero oxygen solution).
Manual salinity and barometric pressure compensation.
selectable readout in ppm, mg/L or percentage saturation.
 

General Features:
Automatic Temperature Compensation  (ATC)  ensures highlyaccurate measurements over the entire range.
Calibration due reminder helps you to calibrate meter regularly.
Automatic endpoint function freezes stable measured values.
Help message as a operational guide to helps you understandhow to use meter.
Setup menu lets user customize the calibration points, stabilityconditions, date and time to meet  measurement requirements.
Expanded memory stores and recalls up to 500 readings.
Built-in real-time clock stamps stored data to meets GLP.
Stored data can be transferred into PC via USB communication interface.

 
Model PH900P
 pH Range -2.000~20.000pH
 pH Accuracy ±0.002pH
 Resolution 0.001, 0.01, 0.1pH, selectable
 Calibration Points Up to 5 points
 Calibration Solutions USA(pH1.68/4.01/7.00/10.01/12.45), NIST(pH1.68/4.01/6.86/9.18/12.45), DIN(pH1.09/4.65/6.79/9.23/12.75)
 Ion Concentration Range 0.001~19999ppm, mg/L, mol/L (Depending on measuring range of Ion selective Electrode)
 Ion Concentration Accuracy ±0.5% Full Scale (Monovalent), ±1% Full Scale (Divalent)
 Calibration Points 2 to 5 points
 Calibration Solutions 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L
 mV Range -1999.9~1999.9mV
 mV Accuracy ±0.2mV
 Resolution 0.1, 1mV, selectable
 Conductivity Range 0~20mS/cm (Max.200mS/cm)
 Conductivity Accuracy ±0.5% Full Scale
 Calibration Points Up to 5 points
 Calibration Solutions 10uS/cm, 84uS/cm, 1413uS/cm, 12.88mS/cm, 111.8mS/cm
 TDS Range 0~10ppt (Max. 20ppt, Depending on factor setting)
 TDS Accuracy ±1% Full Scale
 TDS Factor 0.1~1.0 (Default 0.5)
 Salinity Range 0~10ppt (Max.80ppt)
 Salinity Accuracy ±1% Full Scale
 Salinity Measurement Modes Practical Salinity (PSU) or Seawater Measurement Standard (ppt)
 Resistivity Range 0~100MΩ
 Resistivity Accuracy ±1% Full Scale
 Resolution 0.001, 0.01, 0.1, 1, Automatic
 Temperature Coefficient 0.0~10.0% per °C
 Compensation Modes Linear or Pure Water
 Cell Constant K=0.1, 1, 10, selectable 
 Normalization Temperature 20°C or 25°C
 Dissolved Oxygen Range 0.00~20.00mg/L or ppm
 Resolution 0.01, 0.1mg/L, selectable
 Dissolved Oxygen Accuracy ±0.2mg/L (Calibration and measurement under the conditions of same temperature)
 % Saturation of Oxygen 0.0~200.0%
 Accuracy ±2.0%
 Salinity Correction 0~50g/L
   Pressure Correction 60.0~112.5kPa, 450~850mmHg
   Temperature Range 0~105°C, 32~221°F
   Temperature Accuracy ±0.5°C, ±0.9°F
   Temperature Calibration Manual
   Temperature Compensation 0~100°C, 32~212°F, Automatic or Manual; DO Mode: 0~50°C, 32~122°F, Automatic
   Auto-Off Manual or Automatic (10, 20, 30 minutes after last key pressed, selectable)
   Hold Function Manual or Automatic
   Stability Conditions Low or High, selectable
   Calibration due reminder 0 to 30 days
   Reset Function Yes
   Memory 500 data sets with Date/Time stamping
   Output USB Communication Interface
   Connector BNC and 6-pin connector
   Power Requirements 3×1.5V "AA" Batteries
   LCD Display Custom dual-display LCD (130×110mm)
   Dimensions 170(L)×85(W)×30(H)mm
   Weight 300 grams
 

Company Information
 
BIOBASE group, founded in 1999 and been through high and stable development for 17 years, who is now with revenues of one hundred million, 1200 employees in China and abroad offices, 22.4 acres land industry parks and 20000 square meters production base.

The products we manufacture extends to most of the laboratory items and scientific analytical equipments. Our mission is to enable our customers to make the world healthier, cleaner and safer.

BIOBASE group invests in science and technology research, aim to keep impoving products performance and quality as well as develop new products to meet customers wide requirement.
We satisfy our customers by offering comprehensive support, purchase convenience and we are ready to serves more and more BIOBASE users.

BIOBASE group, serving global distributors and its loyal end users, is becoming one of the most important scientific services supplier.

We have branch office in USA, India and Jordan .
Till now we have exported to more than 200 countries.
Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter
Biobase Laboratory Coloured Portable Multi-Parameter Water Quality Meter


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now