Application: | Laboratory Apparatus |
---|---|
Warranty: | > 1Year |
Detection Method: | Spectrophotometer |
Advantage: | High Sensitive |
Principle: | Classic Spectrometer |
Wavelengths: | UV Spectrometer |
Customization: |
---|
Suppliers with verified business licenses
Model |
BK-UV1000 |
BK-V1000 |
BK-UV1200 |
BK-V1200 |
Optical System |
Single Beam, Grating 1200 lines/mm |
|||
Wavelength Range |
200-1020nm |
325-1000nm |
200-1020nm |
320-1020nm |
Spectral Bandwidth |
4nm |
|||
Wavelength Accuracy |
± 1.8nm (2nm) |
±1.5nm (2nm) |
||
Wavelength Repeatability |
≤ 0.2nm (11D: 1nm; V12, UV12: 0.8nm; UV-11: 1nm) |
|||
Photometric Accuracy |
± 0.002A (0-0.5Abs), ± 0.004A (0.5-1.0Abs), ± 0.4% T (0-100% T) (0.004A, 0.5) |
|||
Photometric Repeatability |
0.001Abs (0-0.5Abs), 0.002Abs (0.5-1.0Abs), ≤ 0.2% T (0-100% T) |
|||
Stray Light |
≤ 0.1 % T @ 360nm; 220nm(0.2) |
≤ 0.05% T @ 360nm (0.2) |
||
Stability |
± 0.001A / h @ 500nm (11D: 0.004, V12, UV12, UV11: 0.002) |
|||
Noise |
± 0.001A |
|||
Display |
128*64 Dots LCD |
|||
Photometric Mode |
T, A, C, E |
|||
Photometric Range |
0-200%T,-0.301-3.0A |
|||
Detector |
Silicon Photodiode |
|||
Light Source |
Deuterium Lamp, Tungsten Lamp |
|||
Input |
Membrane Keypad |
|||
Output |
Parallel Print,USB1.0 |
|||
Compartment |
Solid Sample Holder, Micro Cell Holder, 10-100mm Cell Holder |
Suppliers with verified business licenses