Fully Autoclavable Single Channel Adjustable Volume Pipettes

Product Details
Customization: Available
Brand Name: Biobase
Range: 0.1 To1000
Manufacturer/Factory & Trading Company

360° Virtual Tour

Diamond Member Since 2011

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Importers and Exporters
The supplier has import and export rights
Years of Export Experience
The export experience of the supplier is more than 10 years
Experienced Team
The supplier has 36 foreign trading staff(s) and 33 staff(s) with over 6 years of overseas trading experience
Exhibition Experience
The supplier had participated in offline trade shows, you can check the Audit Report for more information
to see all verified strength labels (34)
  • Fully Autoclavable Single Channel Adjustable Volume Pipettes
  • Fully Autoclavable Single Channel Adjustable Volume Pipettes
  • Fully Autoclavable Single Channel Adjustable Volume Pipettes
Find Similar Products

Basic Info.

Model NO.
Model: 713111010000
Holder
Universal Tip Holder
Transport Package
Standard Packaging
Specification
CE SGS
Trademark
BIOBASE
Origin
China
Production Capacity
1000

Product Description

Fully Autoclavable Single Channel Adjustable Volume Pipettes
*Lightweight, ergonomic, low force design
*Digital display clearly reads volume setting
*The pipettes cover volume range of O.1UI t0 10ml
*Easy to calibrate and maintain with tool supplied
*Design helps avoid repetitive strain injuries
*Calibrated in accordance with IS08655.
Each pipette supplied with individual test certificate
*The low part is available for autoclaving


Model:713111010000
Agronomic in design
Lockable Volume
Tip Ejection Button
Balanced Weight
Lowest Pipetting Force
Universal Tip Holder
 
Agronomic in design
Lockable Volume
Tip Ejection Button
Balanced Weight
Lowest Pipetting Force
Universal Tip Holder
Range : 0.1 t0 2.5 IJI
Range : 1-10 IJI
Range : 2-20 IJI
Range : 10-100 IJI
Range : 100-1000 IJI
Standard
SPECIFICATIONS

MicroPette  Single-Channel Adjustable Volume Pipettes
Model Volume Range  Increment Test Volume(μl) Maximume Permissible
Systematic error 
(Inaccuracy)
Maximume Permissible 
Random error 
(Inaccuracy)
        % μl % μl
71211010000 0.1-2.5μl 0.05μl 2.5 2.50% 0.0625 2.00% 0.05
1.25 3.00% 0.0375 3.00% 0.0375
0.25 12.00% 0.03 6.00% 0.015
71211040000 0.5-10μl 0.1μl 10 1.00% 0.1 0.80% 0.08
5 1.50% 0.075 1.50% 0.075
1 2.50% 0.025 1.50% 0.015
71211050000 2-20μl 0.5μl 20 0.90% 0.18 0.40% 0.08
10 1.20% 0.12 1.00% 0.1
2 3.00% 0.06 2.00% 0.04
71211060000 5-50μl 0.5μl 50 0.60% 0.3 0.30% 0.15
25 0.90% 0.225 0.60% 0.15
5 2.00% 0.1 2.00% 0.1
71211080000 10-100μl 1μl 100 0.80% 0.8 0.15% 0.15
50 1.00% 0.5 0.40% 0.2
10 3.00% 0.3 1.50% 0.15
71211090000 20-200μl 1μl 200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
20 3.00% 0.6 1.00% 0.2
71211110000 50-200μl 1μl 200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
50 1.00% 0.5 0.40% 0.2
71211140000 100-1000μl 5μl 1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
100 2.00% 2 0.70% 0.7
71211160000 200-1000μl 5μl 1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
200 0.90% 1.8 0.30% 0.6
71211170000 1000-5000μl 50μl 5000 0.50% 25 0.15% 7.5
2500 0.60% 15 0.30% 7.5
1000 0.70% 7 0.30% 3
71211330000 2-10ml 0.1ml 10ml 0.60% 60 0.20% 20
5ml 1.20% 60 0.30% 15
2ml 3.00% 60 0.60% 12
 8-channel Adjustable Volume Pipettes
Model Volume Range Increment Test Volume(μl) Maximume Permissible 
Systematic error 
(Inaccuracy)
Maximume Permissible 
Random error 
(Inaccuracy)
        % μl % μl
71212040000 0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
71212060000 5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
71212120000 50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
12-channel Adjustable Volume Pipettes
Model Volume Range Increment Test Volume
(μl)
Maximume Permissible 
Systematic error 
(Inaccuracy)
Maximume Permissible 
Random error 
(Inaccuracy)
        % μl % μl
71213040000 0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
71213060000 5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
71213120000 50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
 Fixed Volume Pipettes
Model Volume Range Increment Test Volume Maximume Permissible 
Systematic error 
(Inaccuracy)
Maximume Permissible 
Random error 
(Inaccuracy)
        % μl % μl
71221180000 5μl - 5μl 1.3% 0.065 1.2% 0.06
71221190000 10μl - 10μl 0.8% 0.08 0.8% 0.08
71221200000 20μl - 20μl 0.6% 0.12 0.5% 0.1
71221210000 25μl - 25μl 0.5% 0.125 0.3% 0.075
71221220000 50μl - 50μl 0.5% 0.25 0.3% 0.15
71221230000 100μl - 100μl 0.5% 0.5 0.3% 0.3
71221240000 200μl - 200μl 0.4% 0.8 0.2% 0.4
71221250000 250μl -- 250μl 0.4% 1.0 0.2% 0.5
71221260000 500μl - 500μl 0.3% 1.5 0.2% 1.0
71221270000 1000μl - 1000μl 0.3% 3.0 0.2% 2.0
71221280000 2000μl - 2000μl 0.3% 6.0 0.15% 3.0
71221290000 5000μl - 5000μl 0.3% 15 0.15% 7.5

Company Information

Fully Autoclavable Single Channel Adjustable Volume Pipettes

Contact me

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More