Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Urine Analyzer |
Still deciding? Get samples of US$ 2566/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
BK-6190 Auto Hematology Analyzer
This 3-part hematology analyzer is used for detecting the parameters of RBC, WBC, HGB and differential counting.
Features:
1.Throughput: 60T/H
2.8 inch touch screen
3.21 parameters + 3 histograms
4.3 counting modes
5.100,000 sample results
6.Support LIS
Model |
BK-6190 |
||
Throughput |
60Tests/hour |
||
Measuring Principle |
Impedance for cell counting Cyanide-free method for HGB |
||
Aperture Diameter |
WBC chamber:100μm,RBC/PLT chamber:70μm |
||
Parameters |
WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR |
||
Histograms |
3 histograms(includingWBC, RBC, PLT histogram) |
||
Sample Volume |
Normal:10μl; Predileted:20μl |
||
Calibration |
Manual & auto calibration |
||
Counting Modes |
Venous, capilary, prediluted |
||
Printer |
Built-in thermal printer,USB to external printer(optional) |
||
Data Input |
8 inch touch screen, keyboard, mouse |
||
Output |
RS232 port, USB port, VGA port. LIS system avaliable |
||
Storage |
100,000 results with histogram |
||
Carry-over rate |
Parameter |
CV |
|
WBC |
≤ 2% |
||
RBC |
≤ 1% |
||
HGB |
≤ 2% |
||
PLT |
≤ 2% |
||
Background |
Parameter |
Background |
|
WBC |
≤ 0.2×10^9/L |
||
RBC |
≤ 0.02×10^12/L |
||
HGB |
≤ 1 g/L |
||
HCT |
≤ 0.5% |
||
PLT |
≤ 10×10^9/L |
||
Precision
|
Parameter |
CV |
Measurement range |
WBC |
≤ 2.0% |
(4.0-15.0×10^9)/L |
|
RBC |
≤ 1.9% |
(3.5-6.0×10^12)/L |
|
HGB |
≤ 1.9% |
(110.0-180.0g)/L |
|
MCV |
≤ 0.4% |
(80.0-110.0)fL |
|
PLT |
≤ 4.0% |
(100.0-500.0×10^9)/L |
|
Linearity |
Parameter |
Measurement range |
CV |
WBC |
(0-5.9×10^9)/L (6.0-99.9×10^9)/L |
(±0.3×10^9)/L ±5% |
|
RBC |
(0-0.99×10^12)/L (1.0-9.99×10^12)/L |
(±0.05×10^12)/L ±5% |
|
HGB |
(0-99.0g)/L (100.0-300.0g)/L |
(±2.0g)/L ±2% |
|
PLT |
(0-99×10^9)/L 100-999×10^9)/L |
(±8.0×10^9)/L ±10% |
|
Power supply |
AC 220±10%,60/50Hz, 110±10%,60Hz |
||
Package Size |
560*470*600mm (W*D*H) |
ReagentPackage Size |
370*320*320mm (W*D*H) |
Gross Weight |
24kg |
ReagentGross Weight |
22kg |