Customization: | Available |
---|---|
After-sales Service: | 24 Hour Online Technical Support |
Warranty: | 1 Year |
Still deciding? Get samples of US$ 700/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
BIOBASE Oil Flashpoint Testing Closed Automatic Flash Point Tester Price
It is used to make determination of the closed cup flash point of the petroleum products.
Features:
* The heating power is adjustable and shown visually by voltmeter.
* The structure is designed small and exquisite with stainless-steel table board
Model |
BK-FP261 |
BK-FP261D |
Ambient Temperature |
15~35ºC |
|
Relative Humidity |
≤85% |
≤85% |
Consumption |
≤650W |
≤600W |
Flash Point Determination |
Mercury-in-glass thermometer. Specifications are as below: Scale -5ºC~110ºC, division 0.5ºC Scale 20ºC~150ºC, division 1ºC Scale 90ºC~370ºC, division 2ºC |
Range: Ambient to230ºC Repeatability:≤3ºC Reproducibility: ≤4ºC Accuracy: 0.1ºC |
Oil Cup |
-Inner diameter: 50.7mm~50.8mm -Depth: 55.7mm~56.0mm -The scribed line depth of capacity of testing oil: 33.9mm~34.3mm -Capacity of testing oil: about 70ml |
|
Stirring Rate |
Procedure A: (90~120)RPM, Procedure B: (250±10)RPM |
|
Igniting Device
|
Igniting source: gas(or other civilian fuels, the same below) |
Electric ignition. Diameter of igniter:0.7mm~0.8mm |
Heating Device |
The heating power is continuously adjustable from 0 W to 600W |
|
Heating Rate |
(1~12)ºC/min; Controllable and adjustable |
Procedure A: (5~6)ºC/min, Procedure B: (1~1.5)ºC/min ProcedureC:(3±0.5)ºC/min |
Power Supply |
AC220V±10%, 50Hz |
|
Dimension(W*D*H) |
370*320*300mm |
510*320*330mm |
Packing Size (W*D*H) |
460*400*410mm |
690*500*600mm |
Gross Weight |
13kg |
33kg |
Factory Views:
Founded in 1999, BIOBASE Group having over 2000 employees (till August, 2018), BIOBASE Group is a new high-tech enterprise focusing on development, production and management of scientific equipment, In Vitro Diagnosis (IVD) instrument and reagent.