• Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)
  • Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)
  • Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)
  • Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)
  • Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)
  • Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)

Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)

Customized: Non-Customized
Structure: Portable
Material: Plastic
Certification: CE, ISO
Application: School, Hospital, Lab
Brand Name: Biobase
Customization:
Diamond Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company

Basic Info.

Model NO.
Single and Multi-Channel Micropipette
Holder
Standard
Range
0.05UL to 0.1 Ml
Channel
One to Twelve
Transport Package
Standard Packaging
Specification
CE SGS
Trademark
BIOBASE
Origin
China
Production Capacity
1000

Product Description

DESCRIPTIONS

All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.

Fully autoclavable
The MicroPette plus pipettes can be fully autoclaved and withstood steam sterilizes at 121ºC, 1 atm for 20 minutes. The pipettes can be autoclaved without special preparation. After autoclaving the pipette must be cooled down and left to dry over 12 hours.

It is recommended to check the performance of the pipette after each autoclaving. It is also recommended to grease the piston and seal of the pipette after each 10 autoclaving.

Features for MicroPette plus
- Fully autoclavable
- Lightweight, ergonomic design,easy to press
- Digital display 
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain 

Features for multi-channel MicroPette plus
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification

- Can be used with universal style pipette tips


SPECIFICATIONS
Single-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume(μl) Maximume Permissible Systematic error 
(Inaccuracy)
Maximume Permissible Random error
   (Inaccuracy)
      % μl % μl
0.1-2.5μl 0.05μl 2.5 2.50% 0.0625 2.00% 0.05
1.25 3.00% 0.0375 3.00% 0.0375
0.25 12.00% 0.03 6.00% 0.015
0.5-10μl 0.1μl 10 1.00% 0.1 0.80% 0.08
5 1.50% 0.075 1.50% 0.075
1 2.50% 0.025 1.50% 0.015
2-20μl 0.5μl 20 0.90% 0.18 0.40% 0.08
10 1.20% 0.12 1.00% 0.1
2 3.00% 0.06 2.00% 0.04
5-50μl 0.5μl 50 0.60% 0.3 0.30% 0.15
25 0.90% 0.225 0.60% 0.15
5 2.00% 0.1 2.00% 0.1
10-100μl 1μl 100 0.80% 0.8 0.15% 0.15
50 1.00% 0.5 0.40% 0.2
10 3.00% 0.3 1.50% 0.15
20-200μl 1μl 200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
20 3.00% 0.6 1.00% 0.2
50-200μl 1μl 200 0.60% 1.2 0.15% 0.3
100 0.80% 0.8 0.30% 0.3
50 1.00% 0.5 0.40% 0.2
100-1000μl 5μl 1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
100 2.00% 2 0.70% 0.7
200-1000μl 5μl 1000 0.60% 6 0.20% 2
500 0.70% 3.5 0.25% 1.25
200 0.90% 1.8 0.30% 0.6
1000-5000μl 50μl 5000 0.50% 25 0.15% 7.5
2500 0.60% 15 0.30% 7.5
1000 0.70% 7 0.30% 3
2-10ml 0.1ml 10ml 0.60% 60 0.20% 20
5ml 1.20% 60 0.30% 15
2ml 3.00% 60 0.60% 12


 
8-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume
(μl)
Maximume Permissible Systematic error
   (Inaccuracy)
Maximume Permissible Random error
   (Inaccuracy)
      % μl % μl
0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
12-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume(μl) Maximume Permissible Systematic error
   (Inaccuracy)
Maximume Permissible Random error
  (Inaccuracy)
      % μl % μl
0.5-10μl 0.1μl 10 1.50% 0.15 1.50% 0.15
5 2.50% 0.125 2.50% 0.125
1 4.00% 0.04 4.00% 0.04
5-50μl 0.5μl 50 1.00% 0.5 0.50% 0.25
25 1.50% 0.375 1.00% 0.25
5 3.00% 0.15 2.00% 0.1
50-300μl 5μl 300 0.70% 2.1 0.25% 0.75
150 1.00% 1.5 0.50% 0.75
50 1.50% 0.75 0.80% 0.4
Fixed Volume Pipettes
Volume Range Increment Test Volume(μl) Maximume Permissible Systematic error
   (Inaccuracy)
Maximume Permissible Random error
   (Inaccuracy)
      % μl % μl
5μl - 5μl 1.3% 0.065 1.2% 0.06
10μl - 10μl 0.8% 0.08 0.8% 0.08
20μl - 20μl 0.6% 0.12 0.5% 0.1
25μl - 25μl 0.5% 0.125 0.3% 0.075
50μl - 50μl 0.5% 0.25 0.3% 0.15
100μl - 100μl 0.5% 0.5 0.3% 0.3
200μl - 200μl 0.4% 0.8 0.2% 0.4
250μl -- 250μl 0.4% 1.0 0.2% 0.5
500μl - 500μl 0.3% 1.5 0.2% 1.0
1000μl - 1000μl 0.3% 3.0 0.2% 2.0
2000μl - 2000μl 0.3% 6.0 0.15% 3.0
5000μl - 5000μl 0.3% 15 0.15% 7.5


Micropette Plus Autoclavable Pipette-with Different Channels (Adjustable/Fixed Volume)

 

Send your message to this supplier

*From:
*To:
avatar Ms. Anna
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Product Groups

Diamond Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company
Number of Employees
730
Year of Establishment
2007-07-18