Single-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||
% | μl | % | μl | |||
0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | ||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | ||
0.5-10μl | 0.1μl | 10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | ||
1 | 2.50% | 0.025 | 1.50% | 0.015 | ||
2-20μl | 0.5μl | 20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | ||
2 | 3.00% | 0.06 | 2.00% | 0.04 | ||
5-50μl | 0.5μl | 50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | ||
5 | 2.00% | 0.1 | 2.00% | 0.1 | ||
10-100μl | 1μl | 100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
10 | 3.00% | 0.3 | 1.50% | 0.15 | ||
20-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
20 | 3.00% | 0.6 | 1.00% | 0.2 | ||
50-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
100-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
100 | 2.00% | 2 | 0.70% | 0.7 | ||
200-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
200 | 0.90% | 1.8 | 0.30% | 0.6 | ||
1000-5000μl | 50μl | 5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | ||
1000 | 0.70% | 7 | 0.30% | 3 | ||
2-10ml | 0.1ml | 10ml | 0.60% | 60 | 0.20% | 20 |
5ml | 1.20% | 60 | 0.30% | 15 | ||
2ml | 3.00% | 60 | 0.60% | 12 |
8-channel Adjustable Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% | μl | % | μl | |||||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 | ||
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 | ||
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 | ||
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 | ||||
12-channel Adjustable Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% | μl | % | μl | |||||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 | ||
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 | ||
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 | ||
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 | ||||
Fixed Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) |
Maximume Permissible Systematic error (Inaccuracy) |
Maximume Permissible Random error (Inaccuracy) |
||||
% | μl | % | μl | |||||
5μl | - | 5μl | 1.3% | 0.065 | 1.2% | 0.06 | ||
10μl | - | 10μl | 0.8% | 0.08 | 0.8% | 0.08 | ||
20μl | - | 20μl | 0.6% | 0.12 | 0.5% | 0.1 | ||
25μl | - | 25μl | 0.5% | 0.125 | 0.3% | 0.075 | ||
50μl | - | 50μl | 0.5% | 0.25 | 0.3% | 0.15 | ||
100μl | - | 100μl | 0.5% | 0.5 | 0.3% | 0.3 | ||
200μl | - | 200μl | 0.4% | 0.8 | 0.2% | 0.4 | ||
250μl | -- | 250μl | 0.4% | 1.0 | 0.2% | 0.5 | ||
500μl | - | 500μl | 0.3% | 1.5 | 0.2% | 1.0 | ||
1000μl | - | 1000μl | 0.3% | 3.0 | 0.2% | 2.0 | ||
2000μl | - | 2000μl | 0.3% | 6.0 | 0.15% | 3.0 | ||